Omega Engineering Oxygen Equipment Dissolved Oxygen System User Manual

Users Guide  
Shop online at  
omega.com  
For latest product manuals:  
omegamanual.info  
DOCN600  
Dissolved Oxygen System  
 
TABLE OF CONTENTS  
GENERAL INFORMATION  
1
2
SPECIFICATIONS  
INSTALLATION  
4
DESCRIPTION OF FUNCTIONS  
START-UP AND OPERATION  
OPERATING HINTS  
8
13  
17  
18  
20  
21  
UTILITY MENU  
DIAGNOSTICS  
TROUBLESHOOTING AND SERVICE  
 
MODEL DOCN600 DISSOLVED OXYGEN ANALYZERS  
AND MODEL DOE-601 DISSOLVED OXYGEN SENSOR  
INSTRUCTION MANUAL  
1.0  
GENERAL INFORMATION  
The DOCN600 Dissolved oxygen analyzer is a versatile industrial microprocessor based instrument.  
Outputs are programmed through the menu with push buttons on the face of the instrument.  
Calibration is also accomplished through the front panel menu. The instrument must be used in  
conjunction with the DOE-601 dissolved oxygen sensor.  
The instrument provides non-isolated 0-5 Vdc, 0-1 mA and isolated 4-20 mA analog outputs. The  
analog outputs may be programmed to represent any segment of the measuring scale.  
A membrane perforation alarm is provided which gives early warning of a membrane failure.  
The instrument can also be used to take a temperature reading and one of the non-isolated analog  
outputs can be used to give a continuous temperature signal.  
The integrity of the system is ensured with a watch-dog timer and system alarm. A password  
feature protects the stored values.  
The instrument is housed in a NEMA 4X enclosure. The standard unit is provided with mounting  
hardware for surface mount applications. Both panel and pipe mount kits can be ordered  
separately.  
Page 1  
 
2.0  
SPECIFICATIONS  
SAFETY & SECURITY:  
Non-volatile memory (EEPROM)  
Passcode protected if selected  
DISPLAY:  
3 1/2 digit LED, 1/2" high digits  
Watch-dog timer monitors microprocessor  
Instrument automatically returns to on-line operation if  
accidentally left in menu mode  
MEASURING RANGES:  
D.O.: 0-20.00 ppm and 0-100% saturation, switch  
selectable  
(This feature may be held disabled if desired)  
Temperature: 0-40°C (32 to 104°F)  
ACCURACY: % Saturation:  
± 1%  
± 0.1  
± 0.2°C  
POWER REQUIREMENTS:  
98-132 Vac, 50/60 Hz (less than 10 VA)  
Optional: 196-264 Vac, 50/60 Hz  
ppm:  
Temp:  
STABILITY: 0.1% of span per 24 hrs., non-cumulative  
AMBIENT CONDITIONS:  
-30 to 50°C (-22 to 122°F)  
0 to 90% R.H. non-condensing  
RESOLUTION: % Saturation:  
0.1  
ppm:  
Temp:  
0.01  
0.1°C  
ANALOG OUTPUTS:  
Non-isolated 0-1 mA, 100 ohms maximum load  
Non-isolated 0-5 Vdc, 1000 ohms minimum load  
Isolated 4-20 mA, 800 ohms maximum load  
The isolated output is isolated from the input, ground  
line power and all other outputs  
REPEATABILITY: 0.1% of span or better  
RESPONSE TIME: 20 seconds to 90% of value upon  
step change at 20°C  
Range Expand: The 4-20 mA analog output can be  
made to represent any segment of the measuring  
scale. Minimum segment is 10% of full scale.  
Output Hold: The analog outputs are automatically  
placed on hold during calibration or other setup  
operations.  
Temperature Output: The 0-5 Vdc output can be  
programmed to follow either the process temperature  
or dissolved oxygen.  
ENCLOSURE: NEMA 4X fiberglass reinforced polyester  
enclosure with four 1/2” conduit holes and mounting  
feet for surface mount  
MOUNTING CONFIGURATIONS:  
Standard is surface mount  
Optional panel mount hardware. Part No. DOCN600-PM  
Optional pipe mount hardware. Part No. DOCN600-PIPE  
NET WEIGHT: 3 1/2 lbs. (1.6 kg)  
TEMPERATURE COMPENSATION:  
Automatic 0 to 100°C (32 to 212°F)  
ALTITUDE COMPENSATION:  
Automatic 0 to 40°C (32 to 104°F)  
DIAGNOSTICS:  
When system error is indicated, call STATUS to obtain  
condition code. On-board simulated input to aid in  
troubleshooting to obtain condition code.  
MEMBRANE ALARM:  
A relay activates if membrane is perforated.  
TEST:  
Display value and analog outputs can be set manually  
to any value for testing and diagnostic purposes. This  
feature allows the outputs to be tested independently  
of process value.  
Page 2  
 
MODEL DOCN600 ADDITIONAL FEATURES  
CONTROL RELAYS:  
Two relays can be independently set for operation in  
response to rising or falling value and for fail-safe  
operation. Deadband (hysteresis) independently  
adjustable.  
Rating: 5A 115/230 Vac, 5A 30 Vdc. SPDT  
ALARM RELAY:  
High-Low with fixed deadband of 2% of full scale.  
Normal or fail-safe operation.  
Rating: 5A 115/230 Vac, 5A 30 Vdc. SPDT  
RELAY INDICATORS:  
Three LEDs indicate status or the three relays  
DOE-601 SENSOR  
TEMPERATURE RANGE:  
0-40°C (32 to 104°F)  
WETTED MATERIALS:  
PVC and Teflon F.E.P.  
FAILURE ALARM:  
5A SPDF relay closes if membrane is punctured. LED  
on panel is illuminated  
CABLE:  
Integral 8 m (26 ft.) Maximum distance to analyzer  
300 m (1000 ft.)  
REPLACEMENT CARTRIDGE:  
Part No. DOE-601-SC. Hermetically sealed.  
MOUNTING HARDWARE  
FLOTATION MOUNTING:  
Model DOE-600-BFMK consists of ball float, extension  
pipe, swivel bracket, junction box and 50 feet of  
interconnect cable  
SUBMERSION MOUNTING:  
Model DOE-600-SMK consists of 5 ft. PVC pipe,  
junction box and 50 feet of interconnect cable  
Page 3  
 
3.0  
INSTALLATION  
3.1 Location  
3.1.1 Locate the instrument within the reach of the cable provided for the DOE-601  
sensor.  
3.1.2 Select an installation site which is:  
• free of mechanical vibrations  
• reasonably clean and dry  
• protected from falling corrosive fluids within the ambient temperature and humidity  
specifications  
• remote from high voltage relay and power switches  
3.2 Type of Mounting  
3.2.1 If the instrument is to be pipe or panel mounted a special hardware kit will be  
required. For panel mount order part number DOCN600-PM. For pipe mount order  
part number DOCN600-PIPE. Instructions for both types of mounting will be  
included with the kits.  
3.2.2 For surface mounting, four feet brackets, together with fastening screws, are  
provided with the instrument. These should be fastened to the back of the  
instrument and then it may be screwed or bolted in the selected location.  
3.3 Conduit Connections  
3.3.1 Four 1/2" conduit holes are provided in the bottom of the enclosure. One of these is  
fitted with an approved water tight plug. To maintain NEMA 4 integrity, approved  
conduit hubs must be used to connect conduit. The hubs must be connected to the  
conduit before being connected to the enclosure. Any unused conduit holes must  
be closed with water tight plugs or connectors.  
3.3.2 For convenience of internal connections the right conduit hole (viewed from the  
front) should be used for power connection; the next hole to the left for relay  
outputs; the next hole to the left for analog outputs and finally the fourth hole for  
sensor input.  
3.4 Electrical Connections  
CAUTION: The instrument operates from line voltage. This constitutes a possible shock hazard.  
Ensure that line power is removed before attempting connections. Note: A separate  
source of line power may have been connected to the floating relay contacts.  
3.4.1 To access the terminal strips open the door of the instrument and then unscrew the  
captive retaining screw near the upper right hand corner of the panel. Now swing  
open the panel to reveal the terminal strip on the power supply circuit board and the  
smaller terminal strip on the back of the swing-out board.  
Page 4  
 
3.5 Sensor Installation, Flotation Mounting  
3.5.1 Refer to diagram below and proceed as follows:  
a) Fasten elbow to 90-inch long pipe  
NOTE: Thread sealant (Teflon tape) is recommended on all mounting hardware  
and sensor threads to avoid leaks  
b) Fasten adapter pipe with union to elbow  
c) Insert adapter pipe with union into flotation ball and tighten union lock ring to  
secure float assembly  
Page 5  
 
d) The sensor may be attached to float assembly in one of two ways:  
1. With sensor EXTENDED FROM BOTTOM of flotation ball:  
Insert 12-inch long extender pipe into the flotation ball and fasten it to the  
end of adapter pipe. This will extend the sensor’s membrane 12 inches  
below the flotation ball. NOTE: Extender pipe supplied only when ordered.  
2. With the sensor membrane FLUSH WITH BOTTOM of flotation ball:  
Disregard using extender pipe and proceed with Step e).  
e) Route sensor cable through assembled pipe sections, starting from bottom  
section, and fasten sensor to pipe.  
CAUTION: Do not remove measuring cell cartridge from hermetically sealed  
package until it is to be used in step f). (Membrane will dry out and  
damage measuring cell).  
f) Install measuring cell cartridge. Unfasten union nut from sensor and make sure  
to remove any moisture from cavity area and brass contact strips. Insert  
cartridge in its correctly “keyed” position. Fasten union not tightly so that O-ring  
compresses to create an effective water-tight seal.  
g) Calibrate system in accordance with the procedure described in Section 5.3  
before placing sensor into operation.  
Page 6  
 
h) Install pivot mounting bracket and mount junction box on the bracket. Insert  
pipe/sensor assembly through the bracket to the desired position. Tighten pipe  
locking screw.  
3.6 Sensor Installation, Submersion Mounting  
3.6.1 a) Refer to diagram below and proceed as follows:  
b) Route sensor cable through pipe and fasten sensor to pipe. The use of thread  
sealant (Teflon tape) on all mounting hardware and sensor threads is  
recommended to avoid leaks.  
CAUTION: Do not remove measuring cell cartridge from hermetically sealed  
package until it is to be used in step b). (Membrane will dry out and  
damage measuring cell).  
c) Install measuring cell cartridge. Unfasten union nut from sensor and make sure  
to remove any moisture from cavity area and brass contact strips. Insert  
cartridge in its correctly keyed position. Fasten union tightly so the O-ring  
compresses to create an effective water-tight seal.  
d) Mount the junction box within reach of the sensor cable.  
e) Calibrate system in accordance with the procedure described in Section 5.3  
before placing sensor into operation.  
3.6.2 The terminal strip on the power supply board at the back of the instrument is  
labeled for power supply, relay outputs and analog outputs. Connect wiring in  
accordance with this labeling. CAUTION: Connecting the line voltage power  
supply to incorrect terminals may cause serious damage.  
3.7 Probe Connections  
3.7.1 The interconnect cable as supplied will normally be connected to the sensor cable  
in the junction box. If this is not the case, connect the six wires and the shield of the  
extension cable to the numbered wires in the junction box as shown below:  
Extension Cable  
Junction Box  
Blue  
Red  
1
2
Black  
Green  
White  
Yellow  
Shield  
3
4
5
6
Unnumbered (Green/Yellow)  
3.7.2 Connect the six wires of the interconnect cable to the terminal strip on the swing-out  
board being sure to match colors as printed on the board. Connect the shield to the  
terminal post located above the terminal strip.  
3.7.3 Running the sensor and interconnect cable in 1/2” metal conduit for protection  
against moisture and mechanical damage is recommended. Do not run power or  
control wiring in the same conduit (“electrical noise” may interfere with sensor  
signal).  
Page 7  
 
4.0  
DESCRIPTION OF FUNCTIONS  
4.1 Overview  
4.1.1 The DOCN600 is a microprocessor based dissolved oxygen analyzers are designed  
for industrial applications. They operate in conjunction with the DOE-601 sensor.  
The software in the unit makes the instrument very easy to operate and maintain.  
4.1.2 The outputs include voltage-free relay contacts and industry standard analog  
transmission signals. A relay to indicate a perforated membrane is provided. The  
analog output signals transmit low power signals to peripherals such as data  
recorders or control systems.  
4.1.3 The software is designed for ease of operation. It uses a simple menu with all items  
indicated on the panel. The user interface consists of six buttons on the front panel.  
The buttons are scanned and responded continuously. In addition, an extensive  
system checking for values and parameters is performed by the software. All of the  
operating parameters are stored in non-volatile memory, without the need of a  
battery.  
4.1.4 The DOCN600 operates like a normal analog converter with a number of additional  
functions made possible by the microprocessor in the instrument.  
Page 8  
 
Some of these are:  
• Recall and easy adjustment of relay and output parameters  
• Push button calibration  
• A HOLD function for outputs  
• Continuous sensor check during measurement  
• Continuous self check and watch-dog timer to ensure correct operation  
• Password protection of stored values  
• Temperature Output  
• Simulated input for testing  
4.2 Calibration  
4.2.1 The DOCN600 is calibrated at the factory, with the sensor purchased with the  
instrument. However, the system should be field calibrated when installed and from  
time to time thereafter. The calibration procedure is given in Section 5.3.  
4.3 Temperature Compensation  
4.3.1 A temperature sensor in the DOE-601 sensor provides the means of continually  
compensating the D.O. reading for variation in the temperature of the process.  
4.4 Relay Outputs  
4.4.1 Three SPDT relays are provided. The normally open contacts, NO, are open when  
no alarm or control is active.  
4.4.2 The two control relay can be programmed to close on either rising or falling D.O.  
They can be set to close at any point on the scale. The deadband, sometimes  
termed hysteresis, defines the point at which the relays open. See Sections 4.8,  
5.8, and 5.9.  
4.4.3 The third relay functions as a high-low alarm. It has two programmable setpoints  
(high and low). The deadband is fixed at 2% of full span. This relay is also  
activated in the event of a perforated membrane. (See Section 4.5) The alarm  
contact can be programmed to also signal memory loss in the DOCN600 controller.  
See Section 4.8, 5.12, and 5.13.  
4.4.4 All relays can be programmed for "fail-safe" operation which reverses the normal  
operation of the relay. In this mode the relays will fail-safe (transfer) in the event of  
power failure.  
4.5 Membrane Alarm Relay  
4.5.1 A single pole double throw relay is provided to give an output in the event of a  
perforated membrane.  
4.5.2 This relay can be programmed to also signal memory loss. See Section 4.8.  
4.5.3 When this relay is activated the contacts will close and the FAIL LED will be  
indicated on the panel.  
Page 9  
 
4.6 Analog Outputs  
4.6.1 The analog output signals consist of a non-isolated 0-1 mA, 0-5 Vdc, and isolated 4-  
20 mA signals.  
4.6.2 From the factory all of the analog outputs have a linear range corresponding to the  
full range of the instrument. The 4-20 mA output can be programmed to another  
linear range by entering two values:  
• Output High: This is the D.O. value at which you wish to have 100% output.  
• Output Low: This is the D.O. value at which you wish to have 0% output.  
4.6.3 When programming, you must ensure that the output range has a span of at least  
10% of the full range of the instrument. For applications that require the output to  
decrease as the process value increases, i.e. an inverted output, the Output High  
value will be less than the Output Low. See Section 5.4.  
4.6.4 The instrument may be used to measure the temperature of the process either in  
Celsius or in Fahrenheit. The DOCN600 does not control temperature but the 0-5  
Vdc and 0-1 mA analog outputs can be dedicated to follow the process  
temperature. See Sections 5.6.  
4.7 Operation Menu  
4.7.1 The operation menu allows the user to recall and to adjust the parameters, required  
by the analyzer functions.  
4.7.2 When the analyzer is powered up, the program will display D.O. readings. None of  
the LEDs in the operation menu will be illuminated.  
4.7.3 Five buttons on the panel are used to operate the menu. Use the CALL button to  
step through the items in the menu, one at a time. The red LEDs beside each item  
makes it very easy to follow the menu. The function of the RUN button is to return  
to the on-line D.O. display from anywhere in the menu. The analyzer has a built-in  
timer which, when enabled, by DIP Switch 7 Bank S1 (Section 4.8), returns the  
program to RUN mode if no button has been pressed for 10 minutes. This time-out  
has the same effect as pressing the RUN button.  
Page 10  
 
4.7.4 When in the menu mode, the display initially shows the current value of the  
parameters, such as the Setpoint of the control relay, while putting all of the outputs  
on hold. The two arrow buttons are used to adjust the display value up or down.  
To accept the new value press ENTER twice. While the value on the display is  
being changed, the relay outputs and the analog outputs remain on hold.  
The items that appear in the operation menu are:  
• Password  
• % Sat/ppm  
• Test  
• Calibration  
• Temperature  
• Status  
• Output High  
• Output Low  
• Relay A Setpoint  
• Relay A Deadband  
• Relay B Setpoint  
• Relay B Deadband  
• Alarm Relay High  
• Alarm Relay Low  
4.8 DIP Switches  
4.8.1 The controls which are frequently used in the normal operation of the instrument are  
all accessible on the control panel. Some switches, which are infrequently used are  
located on the back of the swing-out board.  
4.8.2 The DIP switches are scanned on RESET, power-up and every time the instrument  
is taken into the menu mode. Therefore, after changes to the DIP switch settings,  
you must take the unit offline by pressing CALL in order for the instrument to scan  
the new DIP switch values.  
4.8.3 The following table describes the use of the 16 DIP switches:  
DIP Switch  
Bank S1  
Switch  
OFF  
Switch  
ON  
Description of Use  
1
2
3
4
5
6
7
8
Selects temperature unit  
Enables use of Password Feature  
Fail Safe mode for Relay A  
°F  
YES  
NO  
Rising  
YES  
NO  
°C  
NO  
YES  
Falling  
NO  
YES  
YES  
D.O.  
Direction of Control Relay A  
Alarm Relay to Activate for Memory Loss  
Fail Safe mode for Alarm Relay  
Auto Return from menu if no button pressed  
0-5 Vdc / 0-1 mA selector  
NO  
Temperature  
Page 11  
 
Dip Switch  
Bank S2  
1
2
3
4
5
6
7
8
Fail-safe for Relay B  
Direction of Control for Relay B  
Reserved  
Reserved  
Reserved  
Reserved  
Reserved  
Reserved  
NO  
Rising  
YES  
Falling  
4.9 Output Hold  
4.9.1 Output hold, is a function which freezes all output signals at the last value to  
prevent the occurrence of wild distortions during programming and maintenance.  
4.9.2 When the Operation Menu is entered by pressing CALL, the alarm relays and the  
analog outputs are automatically placed on hold and remain on hold until the  
instrument returns to on line. The output hold will remain for a maximum of 10  
minutes after the last button was pressed, if this feature has been enabled. See  
Section 4.8.  
4.10 Parameter and Operation Checking  
4.10.1 The instrument continuously checks all parameters in its memory, while taking D.O.  
readings. When it detects an invalid value, it flashes the LEDs in the operation  
menu to indicate the parameter is at fault. You must then access the operation  
menu to take corrective actions.  
4.11 Simulated Input for Testing  
4.11.1 TEST is function for test and diagnostic purposes. Values can be simulated  
internally and adjusted with the arrow buttons to set display reading and analog  
outputs to any desired value without disrupting instrument calibration.  
4.12 Utility Menu  
4.12.1 The Utility Menu is provided to enable authorized personnel to change the range  
and fine tune the analog . See Section 7.0.  
4.13 Watchdog Timer and Self Diagnostics  
4.13.1 The DOCN600 continuously monitors the condition of all key components of the  
measuring system to ensure that the measurements are reliable. Invalid entries and  
memory loss are indicated on the panel. See Section 8.0.  
Page 12  
 
5.0  
START-UP AND OPERATION  
5.1 Password  
5.1.1 To enter the menu press CALL then PASSWORD will be indicated. With each  
press of CALL you will step through the menu. When the last item is reached the  
menu wraps around to “%sat/ppm”. If you have enabled PASSWORD by placing  
DIP Switch 2 Bank S1, in the off position you must enter the password “6” when  
PASSWORD is indicated if you wish to change any stored value. You do not need  
to enter the password to proceed to other items in the menu for recall of stored  
values.  
If PASSWORD has not been enabled you do not need to enter the password to  
change stored values.  
5.2 Percent Saturation/ppm  
5.2.1 The instrument provides measurement of dissolved oxygen in either percent  
saturation or ppm. In ppm measurement the reading is automatically compensated  
for temperature changes in the medium.  
a) To select units of measurement enter the menu by pressing CALL to indicate  
%sat/ppm.  
b) Press the UP arrow button to select %sat or the DOWN arrow button or select  
ppm.  
c) Press RUN to return to on line or press CALL for another menu selection.  
5.3 Calibration  
5.3.1 To calibrate the instrument you will need a container of fresh clean water. For best  
results a bubbler such as that used in a tropical fish tank should be in the container.  
a) Clean the sensor with a soft cloth and mild soap solution.  
b) Place the sensor over the water container for a few minutes. (There is no need  
to remove it from the float if so installed.)  
c) Press CALL to enter the menu.  
d) If the password function is enabled, enter the password then press CALL to  
illuminate the “%sat/ppm” indicator. If the password function is not enabled,  
“%sat/ppm” will illuminate at the first press of CALL.  
e) Use the UP arrow button to illuminate the %sat LED located to the right of the  
display.  
f) Press CALL twice to illuminate the CALIBRATION indicator.  
g) Use the arrow buttons to make the display read 100.0.  
h) Press ENTER as soon as the reading stabilizes. The display will flash until  
ENTER is pressed again to confirm entry.  
Page 13  
 
i) If you wish to change the measuring unit to ppm, press CALL a number of times  
to return to “%sat/ppm” and change ppm with the down button.  
j) Press RUN to return to on line measurement.  
NOTE: If a major change is necessary to make the display read 100% the display  
will flash when you press ENTER. This is to warn that you may have made  
a mistake such as using “stale” water. If you are confident your procedure  
was correct, press ENTER a second time. Otherwise, repeat the calibration  
procedure by pressing CALL a sufficient number of times to go through the  
menu and return to CALIBRATION.  
5.4 Analog Output Range Expand  
5.4.1 The 4-20 mA isolated analog output may be spread over any section of the scale as  
long as that section is at least 10% of full scale. The best way to describe this  
setup is by example. Suppose you wish the 4-20mA output to span 5 ppm to 10  
ppm. Proceed as follows:  
a) Press CALL as many times as required until OUTPUT HIGH is indicated. Now  
use the arrows to make the meter read 10 ppm. Press ENTER. The display  
will flash until ENTER is pressed again to confirm entry.  
b) Press CALL once to indicate OUTPUT LOW. Now use the arrow buttons to  
make the meter read 5 ppm. Press ENTER twice as above.  
c) The analog output will now be at 4 mA when the D.O. of the solution is 5 ppm  
and will increase to 20 mA when the D.O. of the solution is 10 ppm.  
d) Press RUN to place the instrument on line or press CALL for another menu  
selection.  
e) To invert the output (i.e. decreasing output for increasing process value) simply  
adjust for 4 mA on OUTPUT HIGH and for 20 mA on OUTPUT LOW.  
NOTE: A range expand of less than 10% of full scale is an invalid entry which will  
be indicated by the LED flashing when you return to RUN. To correct,  
return to the menu and correct the output settings.  
5.5 Test  
5.5.1 To test your setup press CALL the number of times required to indicate TEST. Now  
by using the arrow buttons you can “sweep” through the digital display.  
5.5.2 With the instrument in TEST you can check the analog output at any desired  
reading with a meter connected to the appropriate output terminals.  
5.5.3 Another use of this feature is to check the stability of the system. When in TEST,  
use the arrow buttons to select any value, say 10 ppm. Now return online by  
pressing RUN. Return to the meter some hours or days later. Press CALL to  
indicate TEST. If the instrument is still in calibration and if no one has interfered the  
meter should read 10 ppm.  
Page 14  
 
5.6 Temperature  
5.6.1 The temperature of the process can be read at any time by entering the menu and  
calling for TEMPERATURE. Either °C or °F will be indicated depending on the  
position of DIP switch No. 1. Bank S1. See Section 4.8.  
5.6.2 The 0-5 Vdc and 0-1 mA analog outputs can be dedicated to follow the process  
temperature by simply placing DIP switch No. 8 in the off position. The temperature  
span of the output is set to the utility menu. See Section 4.8 and 7.4.  
5.7 Status  
5.7.1 The DOCN600 continuously checks the integrity of all stored data and monitors the  
condition of the measuring system. If a fault is detected, the FAIL LED above the  
display will turn red. The STATUS in the operation menu will provide a numerical  
code, giving a possible cause and a suggested remedy.  
5.7.2 The following table shows the display codes, causes and remedies:  
Code  
Possible Cause  
Normal Condition  
Reserved  
Suggested Remedy  
0
1
2
No action required  
Temperature Sensor off Scale  
Verify process. Check for open or short  
connections.  
3
D.O. reading off Scale  
Memory Loss  
Verify process. Check for open or short  
connections.  
4*  
Call your OMEGA Rep. Or OMEGA  
directly  
5
6
7
Reserved  
Sensor Fault  
Check probe membrane for perforation  
Factory Setting in force, as a  
Perform procedure according to Section 9.2  
result of the ESCAPE procedure  
NOTE: Code 4 could be a serious failure so the alarm relay will energize in addition to the red  
illumination of the FAIL LED, if DIP switch 5 Bank S1 is OFF.  
5.8 Relay A Setpoint  
5.8.1 As shipped from the factory, Relay A is configured to control decreasing D.O.  
However, you may change the direction of control by changing the position of DIP  
switch 4 Bank S1. Please refer to Section 4.8.  
Page 15  
 
5.8.2 The relay setpoint may be at any place on the scale. To establish the setpoint  
proceed as follows:  
a) Press CALL to enter the MENU. If the PASSWORD function is enabled, enter  
the password then press CALL to illuminate the “%sat/ppm” indicator. If the  
PASSWORD function is not enabled, “%sat/ppm” will illuminate at the first press  
of CALL.  
b) Use the down arrow button to illuminate the PPM LED located to the right side  
of the display.  
c) Press CALL the required number of times until RELAY A SETPOINT is  
indicated. Now use the arrows to make the display read the desired value.  
Press ENTER. The display will flash until ENTER is pressed again to confirm  
entry. A LED above the display indicates when Relay A is activated.  
d) Press RUN to place the instrument on line or press CALL for another menu  
selection.  
5.9 Relay A Deadband  
5.9.1 If you have configured the instrument with the DIP switch on the back of the panel  
for falling D.O. you will wish to have the deadband (sometimes called “hysteresis”)  
at a higher value than the setpoint. If you have established a setpoint of 3 ppm you  
may wish the deadband to be between 6 ppm and 3 ppm. Proceed as follows:  
a) Enter the menu and press CALL the required number of times to indicate  
RELAY A DEADBAND. Use the arrow buttons to make the display read the  
desired value (in this example, 6). Press ENTER. The display will flash until  
ENTER is pressed again to confirm the entry.  
b) In this example, after the instrument has been returned to RUN, the relay will be  
energized when the D.O. falls to 3 ppm and will remain energized until the  
reading rises above 6 ppm.  
c) When Relay A is activated a LED above the display will be illuminated.  
d) Press RUN to place the instrument on line or press CALL for another menu  
selection.  
5.10 Relay B Setpoint  
5.10.1 Relay B Setpoint is configured in the same way as Relay A Setpoint. See Section  
5.8. A LED above the display indicates when Relay B is activated.  
5.11 Relay B Deadband  
5.11.1 Relay B Deadband is configured in the same way as Relay A Deadband. See  
Section 5.9.  
NOTE: A deadband setting on the wrong side of the setpoint in an invalid entry  
which will be indicated by the LED flashing when you return to RUN. To  
correct return to the menu and change the deadband setting or, if the  
direction of control is the problem, change the position of the relevant DIP  
switch. See Sections 4.8 and 5.9.  
Page 16  
 
5.12 Alarm High  
5.12.1 The instrument is fitted with a relay, which is set to activate on both high and low  
alarm conditions. The deadband is factory set. To set the ALARM HIGH proceed  
as follows.  
a) Enter the menu and press CALL until ALARM HIGH is indicated. With the  
arrow buttons make the display read the desired alarm value. Press ENTER.  
The display will flash until ENTER is pressed again to confirm the entry.  
b) Press RUN to place the instrument on line, or press CALL, for another menu  
selection.  
5.13 Alarm Low  
5.13.1 a) Press CALL until ALARM LOW is indicated. With the arrows make the meter  
read the desired value. Press ENTER. The display will flash until ENTER is  
pressed a second time to confirm entry.  
b) Press RUN to place the instrument on line, or press CALL, for another menu  
selection.  
NOTE: A low alarm point higher than the high alarm point is an invalid entry, which  
is indicated by the LED flashing when you return to RUN. To correct, return  
to the menu and reset the alarm points.  
6.0  
OPERATING HINTS  
6.1 Cell Care  
6.1.1 If the sensor membrane is exposed to the atmosphere for more than 24 hours, the  
electrolyte inside the membrane can dry out. If this occurs, the cartridge will need  
to be replaced. To avoid membrane dry-out, place the sensor in a container of  
clean drinking water until it can be reinstalled into the aeration basin.  
6.1.2 Sensor cartridges can be stored for up to 5 years if the seal on its storage container  
is not broken and the temperature is between 70°F and 80°F with 50 to 80%  
humidity. Do not store sensor cartridges at temperatures below freezing point  
(32°F). The electrolyte solution will crystallize at 25-30°F.  
6.1.3 Cleaning the sensor membrane before each calibration is recommended. Use soft  
cloth and mild soap solution to remove material from the membrane. When  
cleaning, take care to avoid puncturing the sensor membrane. Should this occur,  
the complete sensor cartridge must be replaced.  
6.2 Calibration  
6.2.1 Depending on the application, system calibration should be performed periodically  
to maintain measurement accuracy. Frequent checks are suggested until the  
operational history indicates the optimum period between checks.  
Page 17  
 
6.3 Reset  
6.3.1 The instrument can be reset without losing calibration or any of the stored values by  
pressing and releasing the S9 reset button (located at the bottom of the board near  
the center) This action is equivalent to turning the power off and on.  
6.4 Output Hold  
6.4.1 It may be useful during some system maintenance procedures to place the relay  
and analog outputs on hold. To accomplish this simply press CALL. To return to  
on line operation press RUN.  
NOTE: To safeguard against the operator forgetting to press RUN the instrument  
will automatically go back on line ten minutes after the last button was  
pressed provided this feature has been enabled by placing DIP switch  
No.7, Bank S1, in the ON position.  
7.0  
UTILITY MENU  
7.1 Utility Menu Functions  
7.1.1 The Utility Menu enables authorized personnel to perform the following:  
• Adjust the temperature output range.  
• Adjust the offset and span of the 0-1 mA/ 0-5 Vdc output  
• Adjust the offset and span of the 4-20 mA output  
• Perform the altitude compensation  
7.2 Access to Utility Menu  
7.2.1 The Utility Menu is protected by password. To access the Utility Menu press and  
hold both RUN and ENTER for five seconds until the PASSWORD LED illuminates.  
The RUN LED will flash to warn that Utility Menu is in use. Now with the arrow keys  
make the display read the password, "7". Press ENTER.  
7.2.2 The Utility Menu is entirely separate from the Operation Menu but uses the same  
LED display. Press CALL to step through the menu items.  
Below is the cross-reference between menu items:  
Display LED  
Meaning in Utility Menu  
Output High  
Output Low  
%Sat/ppm  
Test  
Calibration  
Temperature  
Status  
Temperature Output 100% (5 Vdc)  
Temperature Output 0% (0 Vdc)  
0-1 mA / 0-5 Vdc Output Adjust Low  
0-1 mA / 0-5 Vdc Output Adjust High  
4-20 mA Output Adjust Low  
4-20 mA Output Adjust High  
Altitude Compensation  
Page 18  
 
7.3 Temperature Output  
7.3.1 The 0-5 Vdc and 0-1 mA outputs may be programmed to track the temperature of  
the process. The factory outputs have a linear range corresponding to 0°C to 40°C,  
(or 32°F to 104°F.) Suppose you wish the output to span 10°C to 45°C. Proceed  
as follows:  
a) Enter the Utility Menu as described in 7.2.1.  
b) Press CALL to indicate OUTPUT HIGH, which is the “Temperature output,  
100% point” in the Utility Menu. (See table in 7.2.2). Now use the arrow button  
to show 45.0 on the display. Press ENTER. The display will flash until ENTER  
is pressed again to confirm entry.  
c) Press CALL to indicate OUTPUT LOW, which is the “Temperature output, 0%,  
point” in the Utility Menu. (See table in 7.2.2). Now use the arrow buttons to  
show 10.0 on the display. Press ENTER twice as above.  
d) Press RUN to return to on line or press CALL to proceed to another item in the  
Utility Menu.  
7.4 Adjust 0-1 mA / 0-5 Vdc Output  
7.4.1 It may be desirable to fine tune the 0-1 mA / 0-5 Vdc output to take into account the  
characteristics of your particular loop. The following method involves a high and  
low calibration and requires the use of a digital multi-meter (DVM). Proceed as  
follows ignoring the instrument display:  
a) Turn off the power to the instrument. Connect your DVM to the 0-5 Vdc output  
terminals on the power supply board.  
b) Turn on the power. Enter the Utility Menu as described in 7.2.1.  
c) Press CALL to indicate “%sat/ppm”, which is "0-1 mA/ 0-5 Vdc output adjust,  
low" in the Utility Menu (See table in 7.2.2). Use the arrow keys to make your  
DVM read 1.25V. Press ENTER twice to confirm.  
d) Press CALL to indicate TEST, which is "0-1 mA / 0-5 Vdc Output adjust, high" in  
the Utility Menu (See table in 7.2.2.) Use the arrow keys to make your DVM  
read 3.75V. Press ENTER twice to confirm.  
e) Press RUN to return to on line or press CALL to proceed to another item in the  
Utility Menu.  
7.5 Adjust 4-20 mA Output  
7.5.1 It may be desirable to fine tune the 4-20 mA isolated output to take into account the  
characteristics of your particular loop. Before deciding that the adjustment is  
necessary, be aware of the "Output High" and "Output Low" settings you may have  
programmed as described in Section 5.4. The following method involves a high and  
low calibration and requires the use of a Digital Multimeter (DVM). Proceed as  
follows ignoring the instrument display.  
a) Turn off the power to the instrument. Connect your DVM to the 4-20 mA output  
terminals on the power supply board.  
Page 19  
 
b) Turn on the power. Enter the Utility Menu as described in 7.2.1.  
c) Press CALL to indicate CALIBRATION, which is "4-20 mA Output adjust, low" in  
the Utility Menu (See table in 7.2.2.) Now use the arrow keys to make your  
DVM read 8 mA. Press ENTER twice to confirm.  
d) Press CALL to indicate TEMPERATURE, which is "4-20 mA Output adjust,  
high" in the Utility Menu (See table in 7.2.2). Now use the arrow keys to make  
your DVM read 16 mA. Press ENTER twice to confirm.  
e) Press RUN to return to on line or press CALL to proceed to another item in the  
Utility Menu.  
7.6 Altitude Compensation  
7.6.1 The instrument can be compensated for the altitude at which it is located by the  
following procedure:  
a) Go to the UTILITY MENU as described in 7.2.1.  
b) Press CALL to indicate “STATUS” which is “ALTITUDE COMPENSATION” in  
the Utility Menu.  
c) Adjust the display to your altitude in thousands of feet. For example, 0.34 on  
display corresponds to an altitude of 340 feet. Press ENTER. The display will  
flash until ENTER is pressed again to confirm entry.  
d) The supported altitude range is from –1,000 to 30,000 ft. (-305 to 9,144 m) or –  
1.00 to 30.00 on the display. As shipped from the factory the altitude is set to  
zero. i.e. sea level.  
e) Press RUN to go on line, or CALL to proceed to another Utility Menu item.  
8.0  
DIAGNOSTICS  
8.1 Description  
8.1.1 The DOCN600 has diagnostic features which alerts the operator to invalid entries  
and memory loss. Invalid entries are indicated by the flashing of the appropriate  
menu LED. The flashing will commence after RUN is pressed and will continue  
until the errors are corrected. Memory loss is indicated by the flashing of TEST and  
by the alarm relay if enabled by DIP Switch No. 5 of Bank S1. See Section 4.8.3.  
8.2 Invalid Output  
8.2.1 Invalid output will be indicated if the expanded range is less than 10% of full scale.  
To correct, refer to Section 5.4.  
Page 20  
 
9.0  
TROUBLESHOOTING AND SERVICE  
9.1 Isolate the cause  
9.1.1 When a measurement problem occurs, the first step is to try to isolate the cause. If  
the DOCN600 is powered, go through the menu and check your settings. A  
convenient way to do this is to call TEST. See Section 5.5.  
9.1.2 If your DOCN600 appears dead or intermittent, check the breaker, make sure that  
the instrument is set up for the available line voltage and make sure the line voltage  
is actually available at the terminals. Now measure that sufficient voltage is  
available at all times; it should be 98 Vac to 132 Vac or 187 Vac to 243 Vac  
respectively. Shut line power off, making sure it is off. CAUTION: Power to the  
relays may be supplied from a separate source, shut it off too. Check and if  
necessary replace the internal 0.25A fuse. Push the connector of the ribbon cable  
firmly into its socket. If these steps do not solve the problem it may be necessary to  
replace the power supply board. See Section 9.3.1 below.  
9.1.3 If the process value seems wrong, clean the sensor as described in Section 6.1.3.  
Inspect the sensor, wire, terminal block connections and interconnections.  
Calibrate and resume operation.  
9.1.4 To find out whether the problem is in the sensor, or in the analyzer, use the self-  
testing features. Leave the instrument in RUN mode and proceed as follows:  
a) Move the slide switches S40 and S41 on the back of the swing-out panel from  
position "ON LINE" to position "TEST."  
b) Press CALL to enter the menu and go to “%sat/ppm.”  
c) Set the D.O. simulation DIP switch Bank S42 switch No. 1, On, Switches No. 2,  
3, and 4, Off. The display should show 4 ppm ±5%.  
d) Now proceed to turn each of the other S42 switches On with the remaining  
switches Off.  
The display should read within 5% of 10 ppm with switch 2 On, 16 ppm with  
switch 3 On, and 20 ppm with switch 4 On.  
If this is the case the analyzer is in order and the problem is with the probe.  
e) Return S40 and S41 to “On line” and ensure that all S42 switches are Off.  
9.1.5 a) To check the temperature channel move the slide switches S40 and S41 from  
position “On line” to position “TEST”.  
f) Set the DIP switch 1 Bank S1 On.  
g) Go to the TEMPERATURE menu item.  
d) Set the temperature simulation DIP switch Bank S43 switch No. 1 On, switches  
No. 2 and 3 Off. The display should show 10 degrees Celsius ±5%.  
Page 21  
 
e) Now proceed to turn each of the other S43 switches On with the remaining  
switches Off.  
The display should read within 5% of 25°C with Switch 2 On and 35°C with  
Switch 3 On.  
If this is the case the analyzer is in order and the problem is in the sensor.  
Otherwise the problem is in the analyzer.  
f) Return S40 and S41 to “On line” and ensure that all S43 switches are Off.  
9.2 Escape  
9.2.1 If the instrument appears to be "DEAD", for example not responding to the buttons,  
or not performing on line measurement and control, always try the reset feature  
first, as described in 6.3.1.  
9.2.2 a) The "ESCAPE" procedure is to be used normally at the factory only, when the  
unit is powered with a new MCU. As a result, the internal non-volatile memory  
(EEPROM) is "FORMATTED" and the factory values are loaded into it.  
b) The "ESCAPE" procedure is to be used if a unit is serviced for a new MCU  
insertion, in case where the MCU was not calibrated at OMEGA, or in case of a  
memory loss problem, when so advised by OMEGA service support.  
9.2.3 Before performing the "ESCAPE" procedure it is important to know that this  
procedure provides the option to erase all programmed values and replace them  
with the factory set default values. This is also true for the control setpoints and  
deadbands, then alarm settings and the analog outputs scaling. Also, the  
temperature and the D.O. calibration points will be set to their initial values. This  
means, that the temperature and the D.O. calibration must be performed after an  
“ESCAPE” procedure. After that, all the control, alarm and scaling parameters  
mentioned above must be set to the user values.  
9.2.4 To perform the ""ESCAPE"" proceed as follows.  
a) Turn off the power.  
b) Press and hold the RUN button for about 3 seconds, while turning on the  
power.  
The FAIL LED will turn on and STATUS LED will become red and status code 7  
will be obtained to show that the ESCAPE procedure was just performed.  
9.2.5 After the "ESCAPE" procedure it is necessary to do the following:  
a) Tune the analog outputs and the temperature output span to suit your particular  
application and loop. See Sections 7.3, 7.4, and 7.5.  
b) Calibrate the system, as described in Section 5.3.  
c) Set up the user values for:  
• Output High and Low - See Section 5.4.  
• Control Relay - See Sections 5.8 and 5.9.  
• Alarm Relay - See Sections 5.12 and 5.13.  
Page 22  
 
d) After all the above operations are performed the FAIL LED will turn off and  
STATUS LED will turn green and Status “0” should be obtained.  
e) The unit may be tested now, using the TEST menu item (See Section  
5.5), or the built-in self-testing feature, (See Section 9.1.4)  
f) Bring the unit on line for measurement and control.  
9.3 Printed Circuit Board Replacement  
9.3.1 a) To replace printed circuit boards or relays shut off all power to the DOCN600,  
including any independent power to the relay contacts. Make a record of the  
external wiring, then disconnect the wires. Unplug the ribbon cable connector.  
b) The power supply circuit board is fastened to the back of the enclosure by four  
screws, remove the screws and the board is free. Reverse the procedure to  
mount a replacement board.  
c) The microprocessor circuit board is located on the swing-out assembly behind  
the door. Swing the assembly out, lift it up to unseat the lower hinge pin (the  
upper hinge pin is spring loaded.) The assembly is now free. The circuit board  
is fastened to the front panel by three screws. Remove the screws to release  
the circuit board.  
9.4 Sensor Fault  
9.4.1 If the membrane of the sensor cartridge is pierced, the alarm relay contacts will  
close and the FAIL LED and STATUS LED will be illuminated. However, it may be  
difficult to see the leakage with the naked eye. A slight pressure on the membrane  
with a finger will produce a drop of liquid. The cartridge must be replaced. See  
Section 9.4.3.  
9.4.2 With operating time silver oxide will build up on the anode (silver wire). When the  
anode becomes brown as a result of the deposit, it is time to replace the sensor  
cartridge.  
9.4.3 To replace the sensor cartridge, unscrew the retaining nut, remove and discard the  
spent cartridge and replace with a new cartridge. Part No. DOE-601-SC.  
9.5 Relay Replacement  
9.5.1 The relays are plugged into the power supply board just above the terminals. Be  
sure all the power including independent power to the relay is off. Unplug the relay  
in question and replace with a new relay. Contact OMEGA Engineering for  
assistance.  
Page 23  
 
STATEMENTS OF CONFORMITY FROM THE MANUFACTURER  
U.S.A.  
Canada  
WARNING: This equipment generates, uses, and can  
radiate radio frequency energy and if not installed and  
used in accordance with the instructions manual, may  
cause interference to radio communications. It has  
been tested and found to comply with the limits for a  
Class A computing device pursuant to Subpart J of Part  
15 of FCC Rules, which are designed to provide  
reasonable protection against such interference when  
operated in a commercial environment.  
This digital apparatus does not exceed the Class A  
limits for radio noise emissions from digital apparatus  
set out in the radio interference regulations of the  
Canadian Department of Communications.  
Le presént appareil numérique n’émet pas de bruits  
radioélectriques dépassant les limites applicables aux  
appareils numériques de la classe A prescrites dans le  
Réglement sur le brouillage radioélectrique édicté par  
le ministére des Communications du Canada.  
Page 24  
 
WARRANTY/DISCLAIMER  
OMEGA ENGINEERING, INC. warrants this unit to be free of defects in materials and workmanship for a  
period of 13 months from date of purchase. OMEGA’s WARRANTY adds an additional one (1) month  
grace period to the normal one (1) year product warranty to cover handling and shipping time. This  
ensures that OMEGA’s customers receive maximum coverage on each product.  
If the unit malfunctions, it must be returned to the factory for evaluation. OMEGA’s Customer Service  
Department will issue an Authorized Return (AR) number immediately upon phone or written request.  
Upon examination by OMEGA, if the unit is found to be defective, it will be repaired or replaced at no  
charge. OMEGA’s WARRANTY does not apply to defects resulting from any action of the purchaser,  
including but not limited to mishandling, improper interfacing, operation outside of design limits,  
improper repair, or unauthorized modification. This WARRANTY is VOID if the unit shows evidence of  
having been tampered with or shows evidence of having been damaged as a result of excessive corrosion;  
or current, heat, moisture or vibration; improper specification; misapplication; misuse or other operating  
conditions outside of OMEGA’s control. Components in which wear is not warranted, include but are not  
limited to contact points, fuses, and triacs.  
OMEGA is pleased to offer suggestions on the use of its various products. However,  
OMEGA neither assumes responsibility for any omissions or errors nor assumes liability for any  
damages that result from the use of its products in accordance with information provided by  
OMEGA, either verbal or written. OMEGA warrants only that the parts manufactured by the  
company will be as specified and free of defects. OMEGA MAKES NO OTHER WARRANTIES OR  
REPRESENTATIONS OF ANY KIND WHATSOEVER, EXPRESSED OR IMPLIED, EXCEPT THAT OF  
TITLE, AND ALL IMPLIED WARRANTIES INCLUDING ANY WARRANTY OF MERCHANTABILITY  
AND FITNESS FOR A PARTICULAR PURPOSE ARE HEREBY DISCLAIMED. LIMITATION OF  
LIABILITY: The remedies of purchaser set forth herein are exclusive, and the total liability of  
OMEGA with respect to this order, whether based on contract, warranty, negligence,  
indemnification, strict liability or otherwise, shall not exceed the purchase price of the  
component upon which liability is based. In no event shall OMEGA be liable for  
consequential, incidental or special damages.  
CONDITIONS: Equipment sold by OMEGA is not intended to be used, nor shall it be used: (1) as a “Basic  
Component” under 10 CFR 21 (NRC), used in or with any nuclear installation or activity; or (2) in medical  
applications or used on humans. Should any Product(s) be used in or with any nuclear installation or  
activity, medical application, used on humans, or misused in any way, OMEGA assumes no responsibility  
as set forth in our basic WARRANTY/DISCLAIMER language, and, additionally, purchaser will indemnify  
OMEGA and hold OMEGA harmless from any liability or damage whatsoever arising out of the use of the  
Product(s) in such a manner.  
RETURN REQUESTS/INQUIRIES  
Direct all warranty and repair requests/inquiries to the OMEGA Customer Service Department. BEFORE  
RETURNING ANY PRODUCT(S) TO OMEGA, PURCHASER MUST OBTAIN AN AUTHORIZED RETURN  
(AR) NUMBER FROM OMEGAS CUSTOMER SERVICE DEPARTMENT (IN ORDER TO AVOID  
PROCESSING DELAYS). The assigned AR number should then be marked on the outside of the return  
package and on any correspondence.  
The purchaser is responsible for shipping charges, freight, insurance and proper packaging to prevent  
breakage in transit.  
FOR WARRANTY RETURNS, please have the  
following information available BEFORE  
contacting OMEGA:  
FOR NON-WARRANTY REPAIRS, consult OMEGA  
for current repair charges. Have the following  
information available BEFORE contacting OMEGA:  
1. Purchase Order number under which the product  
was PURCHASED,  
1. Purchase Order number to cover the COST  
of the repair,  
2. Model and serial number of the product under  
warranty, and  
3. Repair instructions and/or specific problems  
relative to the product.  
2. Model and serial number of the product, and  
3. Repair instructions and/or specific problems  
relative to the product.  
OMEGA’s policy is to make running changes, not model changes, whenever an improvement is possible. This affords  
our customers the latest in technology and engineering.  
OMEGA is a registered trademark of OMEGA ENGINEERING, INC.  
© Copyright 2005 OMEGA ENGINEERING, INC. All rights reserved. This document may not be copied, photocopied,  
reproduced, translated, or reduced to any electronic medium or machine-readable form, in whole or in part, without the  
prior written consent of OMEGA ENGINEERING, INC.  
 
Where Do I Find Everything I Need for  
Process Measurement and Control?  
OMEGA…Of Course!  
Shop online at omega.com  
TEMPERATURE  
ߜ
 
ߜ
 
ߜ
 
ߜ
 
ߜ
 
Thermocouple, RTD & Thermistor Probes, Connectors, Panels & Assemblies  
Wire: Thermocouple, RTD & Thermistor  
Calibrators & Ice Point References  
Recorders, Controllers & Process Monitors  
Infrared Pyrometers  
PRESSURE, STRAIN AND FORCE  
ߜ
 
ߜ
 
ߜ
 
ߜ
 
Transducers & Strain Gages  
Load Cells & Pressure Gages  
Displacement Transducers  
Instrumentation & Accessories  
FLOW/LEVEL  
ߜ
 
ߜ
 
ߜ
 
ߜ
 
Rotameters, Gas Mass Flowmeters & Flow Computers  
Air Velocity Indicators  
Turbine/Paddlewheel Systems  
Totalizers & Batch Controllers  
pH/CONDUCTIVITY  
ߜ
 
ߜ
 
ߜ
 
ߜ
 
pH Electrodes, Testers & Accessories  
Benchtop/Laboratory Meters  
Controllers, Calibrators, Simulators & Pumps  
Industrial pH & Conductivity Equipment  
DATA ACQUISITION  
ߜ
 
ߜ
 
ߜ
 
ߜ
 
ߜ
 
Data Acquisition & Engineering Software  
Communications-Based Acquisition Systems  
Plug-in Cards for Apple, IBM & Compatibles  
Datalogging Systems  
Recorders, Printers & Plotters  
HEATERS  
ߜ
 
ߜ
 
ߜ
 
ߜ
 
ߜ
 
Heating Cable  
Cartridge & Strip Heaters  
Immersion & Band Heaters  
Flexible Heaters  
Laboratory Heaters  
ENVIRONMENTAL  
MONITORING AND CONTROL  
ߜ
 
ߜ
 
ߜ
 
ߜ
 
ߜ
 
ߜ
 
Metering & Control Instrumentation  
Refractometers  
Pumps & Tubing  
Air, Soil & Water Monitors  
Industrial Water & Wastewater Treatment  
pH, Conductivity & Dissolved Oxygen Instruments  
M4177/0805  
 

Multiquip Automobile Parts MQ D306HA User Manual
NEC Answering Machine SV8100 User Manual
Nilfisk Advance America Vacuum Cleaner 01610A User Manual
NordicTrack Treadmill NTL12951 User Manual
Oasis Concepts Heating System Diesel and AC Heating System for Recreational Vehicles and Yachts User Manual
OK International Water Dispenser TS250 User Manual
One for All Universal Remote URC 6131 User Manual
Onkyo Stereo System T 4555 User Manual
Panasonic CRT Television CT 32SC15 CT 32SL15 User Manual
Parkside Engraver PGG 15 A1 User Manual